1-866-4 JJSTECH(1-866-455-7832)

Hanna pH Mini Controller, 115V 0.0 - 14.0 pH - BL981411-1

Item #:
BL981411-1
Brand:
Hanna Instruments
Hanna pH Mini Controller, 115V 0.0 - 14.0 pH - BL981411-1
$200.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Hanna pH Mini Controller, 115V 0.0 - 14.0 pH - BL981411-1

Hanna pH Mini Controller, 115V 0.0 - 14.0 pH - BL981411-1

$200.00

Discontinued


BL981411-1 pH Mini Controller, 115V 0.0 - 14.0 pH by Hanna Instruments

The BL 981411 pH controller has been designed for easy, affordable installation in tight spaces, ideal for simple and effective process control. The unit is provided with high impedance pH input and can be used with any pH electrode with standard BNC connector. Measurements are clearly displayed on the LCD, while the status LED indicates operating mode.

BL 981411 is also provided with a dosing relay. Selecting acid dosing, will cause the relay to activate when the measurement is higher than the set point. If the basic dosing is selected, the relay is activated when the reading falls below the set point.

Set point adjustment (from 0 to 14 pH) and calibration procedures are easily performed with trimmers on the front panel. Users can choose from automatic or manual dosing modes with a switch on the front panel. Manual control is particularly useful during maintenance operations, because it permits operators to enable or disable the dosing relay according to need. An overtime control system advises users when the relay is active too long, to help prevent overdosing.

  • Easy to handle
  • Fire retardant casing
  • Selectable overdose protection system
  • Splash resistant cover

Order Information:

BL 981411-0 (12 VDC) and BL 981411-1 (115/230V) are supplied with mounting brackets, transparent cover and instruction manual.

Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy @ 25°C ±0.1 pH
Calibration manual, with offset trimmer
Dosing Relay max 2A (fuse protected), 250 Vac, 30 Vdc
Dosing Selection acid or alkaline
contact open=acid dosage=relay ON if measurement > set point
contact close=alkaline dosage=relay ON if measurement < set point
Setpoint adjustable from 0.0 to 14.0 pH
Overtime adjustable, typically from 5 to approx. 30 minutes
Input Impedance 102 Ohm
Power Supply 1 = 115/230 Vac; 50/60 Hz Power Supply
Dimensions 79 x 49 x 95 mm (3.1 x 1.9 x 3.7’’)
Weight 300g (10.6 oz)