HI8510 pH Analog Indicator with Self Diagnostic Test by Hanna Instruments

HI 8510 is ideal for monitoring pH in process control. It can provide highly accurate pH measurements and display values on the easy to read LCD. BNC input, amplified probe input and input from transmitter are supported.

Designed for easy and fast installation, the HI 8510 is provided with membrane keypads on the front panel, large display, and auto-diagnostic functions to check pH electrode and instrument status. These instruments also provide ±5V power output and input terminals for amplified electrodes.

The front panel is protected behind a transparent splash-proof cover.

  • Inputs: BNC connector, Direct wire connection with amplified probe (supports HI 1006 series probes) and 4-20 mA from a transmitter
  • Extended temperature compensation range: -20 to 100°C
  • Supports potential matching pin connection
  • Both 4-20 and 0-20 mA analog output
  • ±5V outputs for amplified electrodes
  • Backlit LCD
  • Auto-diagnostic tests for pH electrode and instrument status
  • Operational mode LED indicators
  • Designed for easy panel mount installation
  • Transparent splash-proof cover included

Order Information:

The HI 8510 is supplied complete with mounting brackets and instructions.

Range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy @ 20 °C ±0.02 pH (0 to 100 °C)
±0.05 pH (-20 to 0 °C)
±0.5% (input transmitter)
Input high impedance 1012 Ohm; reference and matching pin inputs are available; 4-20 mA
Power Output ±5 Vcc; 150 mA max load for amplified electrodes
Calibration offset: ±2 pH with OFFSET trimmer
slope: 80 to 110% with SLOPE trimmer
Temperature Compensation fixed or automatic with Pt100, from -20 to 100°C (-4 to 212°F)
Recorder Output 0-20 mA or 4-20 mA (isolated)
Backlight continuous on
Power Supply 115 VAC ±10% or 230 VAC ±10%; 50/60 Hz
Enclosure flame retardant ABS body and front panel; transparent splash-proof front cover
Environment -10 to 50°C (14 to 122°F); RH max 95% non-condensing
Panel Cutout 141 x 69 mm (5.6 x 2.7’’)
Weight 1 kg (2.2 lb.)