1-866-4 JJSTECH(1-866-455-7832)

Troubleshooting

Troubleshooting

If a problem occurs, refer to the solutions provided in Table 15. If the problem persists, contact BW Technologies by Honeywell.

Figure 14. Troubleshooting

Problem Possible CauseSolution
The detector does not activate. Depleted batteryCharge the battery. Refer to Charging the Battery.
Damaged or defective detector Contact BW Technologies by Honeywell.
Detector automatically deactivates.Automatic deactivation due to depleted batteryCharge the battery. Refer to Charging the Battery.
Lockout on Self-Test Error is enabled and a sensor(s) has failed the startup self-testRefer to Lockout on Self-Test Error and Replacing a Sensor or Sensor Filter.
The detector enters alarm immediately when activated.Sensor needs to stabilizeUsed sensor: Wait 60 seconds
New sensor: Wait 5 minutes
Low battery alarmCharge the battery. Refer to Charging the Battery.
Sensor(s) requires calibrationCalibrate the sensor(s). Refer to Calibration.
Hazardous environment Leave the area immediately. Deactivate and reactivate the detector in a safe area that is free of hazardous gas.
The activation startup self-test fails General fault Contact BW Technologies by Honeywell.
Sensor errorReplace the sensor. Refer to Replacing a Sensor or Sensor Filter.
Detector does not display normal ambient gas reading after startup tests.Sensors not stabilizedUsed sensor: Wait 60 seconds
New sensor: Wait 5 minutes
Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration.
Target gas is presentDetector is operating properly. Use caution in suspect areas.
Detector does not respond to pushbutton.Battery is depleted Charge the battery. Refer to Charging the Battery.
Detector is performing operations that do not require user inputPushbutton operation restores automatically when the operation ends.
Detector does not accurately measure gas.Sensor(s) requires calibrationCalibrate the sensor(s). Refer to Calibration.
Detector is colder/hotter than gas temperatureAllow the detector to attain ambient temperature before use.
Sensor filter is blockedReplace the sensor filter. Refer to Replacing a Sensor or Sensor Filter.
Detector does not enter alarm.Alarm setpoint(s) are set incorrectlyReset the alarm setpoints. Refer to Resetting Gas Alarm Setpoints.
Alarm setpoint(s) set to zeroReset the alarm setpoints. Refer to Resetting Gas Alarm Setpoints.
Detector is in calibration mode Complete the calibration procedure.
Detector intermittently enters alarm without reason.Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of the target gas Detector is operating normally. Use caution in suspect areas. Check maximum gas exposure reading.
Alarms are set incorrectly Reset the alarm setpoints. Refer to Resetting Gas Alarm Setpoints.
Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration
Missing or faulty sensor(s)Replace the sensor(s). Refer to Resetting Gas Alarm Setpoints.
Battery has been charging for 6 hours. Charging indicator on LCD shows the battery is still charging.Battery is trickle charging Battery is fully charged and is ready for operation.
Battery indicator does not display when charging. Battery is depleted below normal levelsCharge the battery for 8 hours. If the battery indicator does not light after charging, contact BW Technologies by Honeywell.
Pump is not operating. Diffusion cap is off or not attached correctly.Attach diffusion cap. Ensure seal is tight.
There is a blockage in the tubing.Clear the blockage
Particulate and/or moisture filter requires replacementReplace the filter(s). Replacing the Pump Filters.
Features and options not operating as expected.Changes in Fleet Manager IIVerify that the settings in Fleet Manager II are correct