- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Troubleshooting
Troubleshooting
If a problem occurs, refer to the solutions provided in Table 15. If the problem persists, contact BW Technologies by Honeywell.
Figure 14. Troubleshooting
Problem | Possible Cause | Solution |
The detector does not activate. | Depleted battery | Charge the battery. Refer to Charging the Battery. |
Damaged or defective detector | Contact BW Technologies by Honeywell. | |
Detector automatically deactivates. | Automatic deactivation due to depleted battery | Charge the battery. Refer to Charging the Battery. |
Lockout on Self-Test Error is enabled and a sensor(s) has failed the startup self-test | Refer to Lockout on Self-Test Error and Replacing a Sensor or Sensor Filter. | |
The detector enters alarm immediately when activated. | Sensor needs to stabilize | Used sensor: Wait 60 seconds New sensor: Wait 5 minutes |
Low battery alarm | Charge the battery. Refer to Charging the Battery. | |
Sensor(s) requires calibration | Calibrate the sensor(s). Refer to Calibration. | |
Hazardous environment | Leave the area immediately. Deactivate and reactivate the detector in a safe area that is free of hazardous gas. | |
The activation startup self-test fails | General fault | Contact BW Technologies by Honeywell. |
Sensor error | Replace the sensor. Refer to Replacing a Sensor or Sensor Filter. | |
Detector does not display normal ambient gas reading after startup tests. | Sensors not stabilized | Used sensor: Wait 60 seconds New sensor: Wait 5 minutes |
Sensor(s) requires calibration | Calibrate the sensor(s). Refer to Calibration. | |
Target gas is present | Detector is operating properly. Use caution in suspect areas. | |
Detector does not respond to pushbutton. | Battery is depleted | Charge the battery. Refer to Charging the Battery. |
Detector is performing operations that do not require user input | Pushbutton operation restores automatically when the operation ends. | |
Detector does not accurately measure gas. | Sensor(s) requires calibration | Calibrate the sensor(s). Refer to Calibration. |
Detector is colder/hotter than gas temperature | Allow the detector to attain ambient temperature before use. | |
Sensor filter is blocked | Replace the sensor filter. Refer to Replacing a Sensor or Sensor Filter. | |
Detector does not enter alarm. | Alarm setpoint(s) are set incorrectly | Reset the alarm setpoints. Refer to Resetting Gas Alarm Setpoints. |
Alarm setpoint(s) set to zero | Reset the alarm setpoints. Refer to Resetting Gas Alarm Setpoints. | |
Detector is in calibration mode | Complete the calibration procedure. | |
Detector intermittently enters alarm without reason. | Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of the target gas | Detector is operating normally. Use caution in suspect areas. Check maximum gas exposure reading. |
Alarms are set incorrectly | Reset the alarm setpoints. Refer to Resetting Gas Alarm Setpoints. | |
Sensor(s) requires calibration | Calibrate the sensor(s). Refer to Calibration | |
Missing or faulty sensor(s) | Replace the sensor(s). Refer to Resetting Gas Alarm Setpoints. | |
Battery has been charging for 6 hours. Charging indicator on LCD shows the battery is still charging. | Battery is trickle charging | Battery is fully charged and is ready for operation. |
Battery indicator does not display when charging. | Battery is depleted below normal levels | Charge the battery for 8 hours. If the battery indicator does not light after charging, contact BW Technologies by Honeywell. |
Pump is not operating. | Diffusion cap is off or not attached correctly. | Attach diffusion cap. Ensure seal is tight. |
There is a blockage in the tubing. | Clear the blockage | |
Particulate and/or moisture filter requires replacement | Replace the filter(s). Replacing the Pump Filters. | |
Features and options not operating as expected. | Changes in Fleet Manager II | Verify that the settings in Fleet Manager II are correct |