- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Calibration
Calibration
Guidelines
Recommended gas mixture:
CO: 100 ppm balance N2
H2S: 25 ppm balance N2
LEL: 50% LEL or 2.5% by vol. methane balance air
O2: clean air, 18%
- To ensure accurate calibration, use a premium-grade calibration gas. Gases approved by the National Institute of Standards and Technology (NIST) improve the validity of the calibration.
- Do not use a gas cylinder past its expiration date
- Calibrate a new sensor before use. Install the sensor, activate the
detector, and allow the sensor to stabilize before starting calibration
or a bump test.
Used sensor: 60 seconds / new sensor: 5 minutes - Calibrate the detector at least once every 180 days, depending upon use and sensor exposure to poisons and contaminants.
- Calibrate the detector if the ambient gas varies during startup.
- Calibrate only in a safe area that is free of hazardous gas
- Do not calibrate the detector during or immediately after charging is complete.
- The H2S, CO, LEL and O2 sensors can be automatically zeroed each time the detector is activated if the Auto-Zero on Startup option is enabled. Activate the detector in a normal (20.9%/20.8% O2) atmosphere.
- If a certified calibration is required, contact BW Technologies by Honeywell.
- Calibration can be performed using either a 0.5 l/min regulator or a demand flow regulator.
- The maximum hose length for calibration is 3 ft (0.9 m).
Warning
Calibration cylinders that are used with a demand flow regulator must meet the following maximum inlet pressure specifications:
- Disposable cylinders 0-1000 psig/70 bar
- Refillable cylinders 0-3000 psig/70 bar
BW recommends using premium grade calibration gases and cylinders that are certified to National Standards. The calibration gases must meet the accuracy of the detector.
Diagnostics Protection
The detector tests the ambient air (auto-zero) and the test gas that is applied (auto span) to ensure it meets expected values. Auto-zero sets the zero-gas level of the sensor. If the target gas is present, the zero level will be incorrect.
If excessive target gas is present, the sensor(s) will fail and an error message displays.
In auto span, an error message displays if the target gas does not meet the expected values.
Sensor(s) that fail to span retain the previous span value and do not continue the calibration process.
Connecting the Gas Cylinder to the Detector
Refer to the following procedures and Figure 6. to connect the gas cylinder to the detector for calibration.
Note
Read all of the following procedures before beginning calibration.
Caution
The maximum hose length for calibration is 3 ft (0.9 m).
1. Verify the calibration gas being used matches the span concentration value(s) that are set for the detector.
2. Attach a 0.5 l/min regulator or a demand flow regulator to the gas cylinder.
3. Connect the calibration hose to the regulator on the gas cylinder.
4. Begin the calibration procedures. Refer to the Calibration Procedure section.
5. When calibration is complete, disconnect the hose from the detector and the regulator.
6. Ensure the gas cylinder is stored according to the manufacturer