The CX100 is a loop powered conductivity transmitter for use with contacting conductivity sensors. The 4-20mA output is ideal for connecting to a PLC for advanced control systems. Easily power the unit with available two-wire power supply loop. Versatile mounting options for wall or panel mount.

Features:

  • Flexible configurations: Six selectable conductivity ranges
  • Facilitates easy sensor maintenance: Output 'HOLD' for calibration & setup to avoid process disruption
  • Accurate conductivity / TDS measurements: Integrated Automatic Temperature Compensation (ATC)
  • Compatible with a wide variety of sensors: For use with all Sensorex contacting conductivity sensors
  • One year limited warranty: We stand behind all our products with the Sensorex Guarantee
Measurement Range: (R1) 0-19.99 uS/cm, 0.01 uS/cm resolution; (R2) 0-199.9 uS/cm, 0.1 uS/cm resolution; (R3) 0-1999 uS/cm, 1 uS/cm resolution; (R4) 0-10.00 mS/cm, 0.01 mS/cm resolution; (R5) 0-19.99 mS/cm, 0.01 mS/cm resolution; (R6) 0-199.9 mS/cm, 0.1 mS/cm resolution
Cell Constants: k=0.1 (R1-R5), k=1.0 (R1-R6), k=10 (R2-R6)
Measurement Accuracy: +/- 1% full scale range
Temperature Range: 0 °C to 100 °C , 0.1° C resolution, +/- 0.5 °C accuracy
Calibration Mode: Automatic with built-in standard solutions manual cell constant adjustment
Output: Isolated DC 4-20mA
Enclosure: NEMA 4X IP65, with push-button control pad. HWD: 96 mm (3.77") x 96 mm (3.77") x 66mm (5.20"), 1/4 DIN
Mounting Options: Panel or Wall mount
Ambient Operating Temperature: 0° C to 50° C
Display: Dual display custom LCD
Temperature Compensation: Pt100 RTD with automatic compensation or manually configured
Power Requirement: 12-24V DC, maximum 600Ω load
Certifications: CE
Warranty: One year limited warranty