WD-35408-00 CON 600 Portable Waterproof Conductivity Meter with Probe by Oakton

Conductivity, TDS, resistivity, salinity—measure from pure water to seawater!

  • Durable waterproof and dustproof design with IP67 rating – Even with no probes attached
  • Single or multipoint calibration; auto or manual – Fast, trouble-free calibration for more accurate measurements
  • Large backlit graphic display – Multi-line display with electrode status indicator, calibration data, and more
  • User-settable “calibration due” alarm – Out-of-date or unperformed calibrations are now things of the past!
  • Built-in real-time clock – Time-and-date stamping meets Good Laboratory Practice (GLP) standards
  • Advanced data management – Store up to 500 data sets and download via Infrared IrDA wireless interface
  • Set point alarms – Audible warning when readings are outside set points limit
  • Accept 2-cell and 4-cell conductivity probes – Measure over a broad conductivity range up to 500 mS/cm
  • Password protection – Security for calibration and setup menus
  • Rugged rubber boot – Protects meter from bumps and dings (optional; included with kits)
Conductivity 0 to 200.0 mS
TDS 0 to 200 ppt
Temperature –10.0 to 110.0°C (14.0 to 230.0°F)
Resolution 0.05% full-scale; 0.1ºC (0.1ºF)
Accuracy ±1% full-scale; ±0.5ºC (±0.9ºF)
Calibration Manual cal, up to 5 points (1 per range); automatic up to 4 points (84.0 μS/cm, 1413 μS/cm, 12.88 mS/cm, 111.8 mS/cm). Temperature offset 0.1°C (0.1°F) increments.
TDS factor 0.40 to 1.00
Cell constant 0.010 to 10.000
Memory up to 500 sets with GLP date and time
Output infrared, IrDA
Real-time clock time-and-date stamp on calibration and stored data
Temperature normalization 15 or 30ºC
Temperature coefficient linear or pure (610 only)
Temperature compensation automatic or manual (selectable) from 0 to 100°C
Ambient operating temperature 0 to 50°C (32 to 122°F)
Power four 1.5 V AA batteries (included) or optional universal AC adapter, up to 200 hours continuous use
Dimensions Meter: 7.25"L x 3.25"W x 2.25"H (18.4 x 8.3 x 5.7 cm)
Boxed: 9.13"L x 9.13"W x 2.75"H (23 x 23 x 7 cm)
Weight Meter: 1 lb (0.5 kg)