- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Troubleshooting
Troubleshooting
If a problem occurs, refer to the solutions in the Troubleshooting section.
If the problem persists, contact JJS Technical Services
Table 15. Troubleshooting
Problem | Possible Cause | Solution |
Startup | ||
Startup | Depleted battery | Charge battery. Refer to Charging the Battery |
Damaged or defective detector | Contact JJS Technical Services. | |
The detector enters alarm immediately when activated. | Sensor needs to stabilize | Used sensor: wait 60 seconds New sensor: wait 5 minutes (The oxygen sensor in the X3 needs 60 minutes to stabilize) |
Low battery alarm | Charge battery. Refer to Charging the Battery. | |
Detector requires calibration | Calibrate the detector. Refer to Calibration. | |
Hazardous environment | Leave the area immediately. Deactivate and reactivate in a safe area that is free of hazardous gas, in an atmosphere of 20.9% oxygen. | |
The activation self-test fails. | General fault | Contact BW Technologies by Honeywell. |
Sensor failure | Replace the sensor. Refer to Replacing a Sensor or Sensor Filter. | |
Detector automatically deactivates during startup. | Battery power too low to operate | Charge battery. Refer to Charging the Battery. |
Force Calibration When Overdue option is enabled and calibration is not attempted | Calibrate the sensor(s) immediately. Refer to Calibration. | |
Force Bump When Overdue option is enabled and a bump test is not attempted | Bump test the sensor(s) immediately. Refer to Bump Test. | |
Detector Operation | ||
Detector does not display expected gas readings after activation self-test. | Sensor not stabilized | Used sensor: wait 60 seconds New sensor: wait 5 minutes (The oxygen sensor in the X3 needs 60 minutes to stabilize) |
Sensor(s) requires calibration | Calibrate the sensor(s). Refer to Calibration. | |
Target gas is present | Detector is operating properly. Use caution in suspect areas. | |
Detector does not respond to pushbutton. | Battery is depleted | Charge battery. Refer to Charging the Battery. |
Detector is performing operations that do not require user input | Pushbutton operation restores automatically when the operation ends. | |
Detector does not accurately measure gas. | Sensor(s) requires calibration | Calibrate the sensors. Refer to Calibration. |
Detector is colder/hotter than ambient gas | Allow the detector to attain ambient temperature before use. | |
Sensor filter is blocked | Replace the sensor filter. Refer to Replacing a Sensor or Sensor Filter. | |
Detector does not enter into alarm. | Alarm setpoint(s) defined incorrectly | Reset alarm setpoints. Refer to Factory Gas Alarm Setpoints and Sensor Configuration. |
Alarm setpoint(s) set to zero | Reset alarm setpoints. Refer to Factory Gas Alarm Setpoints and Sensor Configuration. | |
Detector is in calibration mode | Complete the calibration procedure. | |
Detector intermittently enters alarm without reason. | Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of the target gas | Detector is operating normally. Use caution in suspect areas. Check the peak (maximum) gas exposure reading. |
Alarm setpoints defined incorrectly | Reset alarm setpoints. Refer to Factory Gas Alarm Setpoints and Sensor Configuration. | |
Detector requires calibration | Calibrate the sensors. Refer to Calibration. | |
Missing or faulty sensor(s) | Replace the sensor. Refer to Replacing a Sensor or Sensor Filter. | |
Features and options are not operating as expected. | Changes have been made in Fleet Manager II | Verify settings in Fleet Manager II are correct. |
Charging | ||
Battery has been charging for 3+ hours (XT model) or 6+ hours (XL-X3 models). The charging indicator on the detector LCD shows the battery is still charging. | Battery is trickle charging | Battery is fully charged and ready for operation. |
Battery indicator does not display when charging. | Detector is depleted below normal levels | Charge the battery for 8 hours. Detector LEDs may light during first 5 hours. This is normal. If the battery indicator does not light after charging for 8 hours, contact BW Technologies by Honeywell. |
When detector is activated after charging, the battery indicator does not display. | Battery is defective | Contact BW Technologies by Honeywell. |