- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Hanna Research Grade Conductivity/TDS Meter with USP EC/TDS/Resistivity/Salinity and Temperature, 230V - HI4321-01
Item #:
HI4321-02
Brand:
Hanna Instruments
Share this product:
HI4321-02 Research Grade Conductivity/TDS Meter with USP EC/TDS/Resistivity/Salinity and Temperature, 230V by Hanna Instruments
HI 4321 is a research-grade EC/Resistivity/TDS/Salinity benchtop meter with a large, color, graphic LCD with backlight, capable of millesimal measuring resolution of conductivity with an extended range from 0.001 µS/cm to 1 S/cm.
Conductivity is fully customizable and include: temperature compensation coefficient, temperature reference, selectable compensation method, linear, natural water and no compensation, adjustable cell constant and TDS factor.
All ranges of conductivity, resistivity and TDS feature autorange or users can select the unit to measure manually. Three salinity scales are available: natural sea water, practical salinity and %.
HI 4321 can be used to perform all 3 stages of USP <645> method required for conductivity measurement of pure and ultra pure water. The instrument provides clear directions on how to perform each testing stage and automatically monitors the temperature, conductivity and stability during testing and determines whether a sample is within USP limits.
This instrument is equipped with auto standard recognition and can support custom calibration solutions. Up to a four point calibration can be obtained for enhanced accuracy over an extended measuring range.
The enhanced warning system is in place to alert users when measuring outside the calibration range. This system will also remind the user when a new calibration is due.
Features also include real-time graphic displays and on-screen GLP data. Fully customizable instrument and parameters via setup screens. On-board contextual help can be accessed from any mode simply by pressing the HELP button.
Up to 10 profiles can be saved and recalled eliminating the need to reconfigure each time when a different electrode is used.
Three selectable logging modes are available: Automatic, Manual and AutoHold logging. Up to 100 logging lots with 10,000 records per lot can be stored for automatic or manual modes along with up to 200 USP reports.
All features are available in English, Italian, Portuguese and Spanish. Connect to a PC via the USB or RS232 ports, with the use of HI 92000 Windows® compatible software.
- EC, resistivity, TDS and salinity ranges
- Extended range from 0.001 µS/cm to 1000 mS/cm
- Ready to perform all three stages of USP <645> method required for conductivity measurement of pure and ultra pure water
- Three salinity scales: practical salinity, natural sea water, percent
- Linear and natural water temperature compensation, no compensation for actual conductivity
- Fully customizable
- Large log memory with different logging methods
Order Information:
HI 4321-01 (115V) and HI 4321-02 (230V) are supplied with HI 76312 conductivity probe, HI 76404N electrode holder, 12 VDC power adapter and instructions.
EC Range | 0.000 to 9.999 µS/cm; 10.00 to 99.99 µS/cm; 100.0 to 999.9 µS/cm; 1.000 to 9.999 mS/cm; 10.00 to 99.99 mS/cm; 100.0 to 1000 mS/cm (actual EC) |
Resistivity Range | 1.0 to 99.9 Ohms x cm; 100 to 999 Ohms x cm; 1.00 to 9.99 kOhms x cm; 10.0 to 99.9 kOhms x cm; 100 to 999 kOhms x cm; 1.00 to 9.99 MOhms x cm; 10.0 to 100.0 MOhms x cm |
TDS Range | 0.000 to 9.999 ppm; 10.00 to 99.99 ppm; 100.0 to 999.9 ppm; 1.000 to 9.999 ppt; 10.00 to 99.99 ppt; 100.0 to 400.0 ppt actual TDS (with 1.00 factor) |
Salinity Range | practical scale: 0.00 to 42.00 psu; natural sea water scale: 0.00 to 80.00 ppt; percent scale: 0.0 to 400.0% |
Temperature Range | -20.0 to 120°C; -4.0 to 248.0°F; 253.15 to 393.15K |
EC Resolution | 0.001 µS/cm; 0.01 µS/cm; 0.1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm |
Resistivity Resolution | 0.1 Ohms x cm; 1 Ohms x cm; 0.01 kOhms x cm; 0.1 kOhms x cm; 1 kOhms x cm; 0.01 MOhms x cm; 0.1 MOhms x cm |
TDS Resolution | 0.001 ppm; 0.01 ppm; 0.1 ppm; 0.001 ppt; 0.01 ppt; 0.1 ppt |
Salinity Resolution | 0.01 for practical scale/natural sea water scale; 0.1% for percent scale |
Temperature Resolution | 0.1°C; 0.1°F; 0.1K |
EC Accuracy | ±1% of reading (±0.01 µS/cm) |
Resistivity Accuracy | ±2% of reading (±1 Ohm x cm) |
TDS Accuracy | ±1% of reading (±0.01 ppm) |
Salinity Accuracy | ±1% of reading |
Temperature Accuracy | ±0.2°C; ±0.4°F; ±0.2K (excluding probe error) |
EC Calibration | auto standard recognition, custom calibration solution/four point calibration |
Salinity Calibration | percent scale—one point (with HI 7037 standard) |
Temperature Calibration | three points |
Temperature Compensation | linear and non-linear (natural water) |
TDS Factor | 0.40 to 1.00 |
EC Probe | HI 76312 platinum, four ring conductivity/TDS probe with internal temperature sensor and 1 m (3.3’) cable (included) |
Temperature Probe | HI 7662-T stainless steel temperature probe with 1 m (3.3’) cable (included) |
Record Logging | 100 lots with 10,000 record/lot |
Interval Logging | settable between one and max log time |
Type Logging | automatic, log on demand, AutoHold |
Replatinization | yes |
Display | 240 x 320 dot-matrix color LCD with on-screen help, graphing, language selection and custom configuration |
PC Connection | RS232, USB |
Power Supply | 12 VDC adapter (included) |
Environment | 0-50ºC (32 to 122ºF) (273 to 323K) RH max 95% non-condensing |
Dimensions | 160 x 231 x 94 mm (6.3 x 9.1 x 3.7”) |
Weight | 1.2 Kg (2.6 lbs.) |