Discontinued


HI2216-01 0.001 Resolution pH/ORP/ISE/°C Benchtop Meter by Hanna Instruments

The HI 2216 is a pH, ORP, ISE meter with five point pH calibration and 0.001 pH resolution.

This instrument provides GLP capabilities to allow for the storage and retrieval of all data regarding pH, ORP, and ISE calibration.

HI 2216 can perform measurements using ORP electrodes in the mV scale and ISE electrodes in the ppm scale through the pH channel input. A relative mV feature is also provided.

  • Up to five point pH calibration with seven standard buffers
  • Up to two point ISE calibration with five standard solutions
  • 0.001 pH resolution
  • Calibration expiration reminder
  • GLP features
  • Automatic Temperature Compensation
  • Manually log up to 200 records and interval log up to 500 records
  • PC interface via USB

Order Information:

HI 2216-01 (115V) and HI 2216-02 (230V) is supplied with HI 1131B pH electrode, HI 7662 temperature probe, HI 76404N electrode holder, HI 70004 pH 4.01 buffer solution sachet, HI 70007 pH 7.01 buffer solution sachet, HI 7071S electrolyte solution (30 mL), HI 700661 cleaning solution sachet, 12 VDC adapter and instructions.

pH Range –2.0 to 16.0 pH; –2.00 to 16.00 pH; –2.000 to 16.000 pH
mV Range ±999.9 mV (ORP); ±2000 mV (ORP)
ISE Range 0.001 to 19990 ppm
Temperature Range –20.0 to 120.0 ºC (-4.0 to 248.0ºF)
pH Resolution 0.1 pH; 0.01 pH; 0.001 pH
mV Resolution 0.1 mV (±999.9 mV); 1 mV (±2000 mV)
ISE Resolution 0.001 (to 1.999 ppm); 0.01 (to 19.99 ppm);
0.1 (to 199.9 ppm); 1 (to 1999 ppm); 10 (to 19990 ppm)
Temperature Resolution 0.1 ºC
pH Accuracy (@ 20ºC/68ºF) ±0.1 pH; ±0.01 pH; ±0.002 pH
mV Accuracy (@ 20ºC/68ºF) ±0.2 mV (±999.9 mV); ±1 mV (±2000 mV)
ISE Accuracy (@ 20ºC/68ºF) ±0.5% FS
Temperature Accuracy (@ 20ºC/68ºF) ±0.2ºC (excluding probe error)
Relative mV Offset ±2000 mV
pH Calibration automatic, up to five point calibration with seven standard buffers available (1.68, 4.01, 6.86, 7.01, 9.18, 10.01, 12.45), and two custom buffers
ISE Calibration automatic, one or two points with five available buffers (0.1, 1, 10, 100, 1000 ppm)
Temperature Compensation manual or automatic (with HI 7662 probe) from –20.0 to 120.0 ºC (-4.0 to 248.0ºF)
pH Electrode HI 1131B glass body pH electrode with BNC connector and 1 m (3.3’) cable (included)
Temperature Probe HI 7662 stainless steel temperature probe with 1 m (3.3’) cable (included)
Input Impedance 1012 ohm
PC Connectivity opto-isolated USB
Data Logging log on demand, 200 records; autologging, 500 records
Logging Interval stability logging (“StAb”) 5, 10, 30 seconds;
1, 2, 5, 10, 15, 30, 60, 120, 180 minutes
Power Supply 12 VDC adapter (included)
Environment 0 to 50°C (32 to 122°F); RH max 95%
Dimensions 235 x 222 x 109 mm (9.2 x 8.7 x 4.3”)
Weight 1.3 Kg (2.9 lb.)