- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
GasAlertMax XT II Replacement Parts and Error Codes
Service Advantages
- Battery is user/distributor serviceable
- Pump is user/distributor serviceable
- Rugged bass fittings and machine screws to prevent stripping
- Armored charger pins
- Inline pump filtration
Error Codes
Code | Meaning | Cause | Recovery |
14 | Memory corruption on configuration options and/or calibration information | This erro has only been seen with power-related issues - either a low battery or having the battery pulled while the unit is saving its settings | Instrument will automatically reset all settings back to factory defaults. To reconfigure the instrument use Fleet Manager II. Calibration is required after this error |
15 | Watchdog error | Code is stuck in an infinite loop and reset itself | None required |
16 | RAM failure | Bad processor | Replace processor |
17 | Program CRC mismatch | Bad code in processo | Reprogram processor - through IR if possible otherwise it may need to be reprogrammed through JTAG interface |
22-28 | Processor errors | Processor anomaly | Replace processor |
32 | DataFlash error | Unable to communicate the DataFlash for logging or retrieving configuration options | Replace DataFlash |
33 | Timer error | Coding error | Report to BW |
34 | SPI error | Likely a DataFlash communication issue | Power-cycle unit. If error continues, try replacing DataFlash |
35 | Clock error | RTC communication issue or the coin cell battery is too low | Replace coin cell or RTC |
36 | TWI error | Coding error | Report to BW |
37 | RDAC error | Unable to communicate with the RDAC | Replace RDAC |
Parts List
Description | Max XT | Max XT II |
O2 sensor | SR-X-MC | SR-X10-C1 |
CO sensor | SR-M-MC | SR-M-MC |
H2S | SR-H-MC | SR-H-MC |
LEL | SR-W-MP75C | SR-W-MP75C |
Sensor Screen (kit of 2) | XT-SS | XT-SS-1 |
Sensor Screen (kit of 10) | XT-SS-K1 | XT-SS-2 |
Main PCB - replacement | XT-MPCB1 | XT-MPCB2 |
Front Shell - Yellow | XT-FC1 | XT-FC1-1 |
Back Shell - Yellow | - | XT-BC1 |
Front Shell - Black | XT-FC1B | XT-FC1B-1 |
Back Shell - Black | - | XT-BC1B |
Replacement Battery | XT-BAT-K1 | XT-BAT-K1 |
Repacement Pump | XT-RPUMP-K1 | XT-RPUMP-K1 |
Hydrophobic filter kit of 5 | XT-RF-H5 | XT-RF-H5 |
Particulate Filter kit of 5 | GA-PFMAX-5 | GA-PFMAX-5 |
Auxilary Filter kit of 5 | M5-AF-K2 | M5-AF-K2 |
Max XT II upgrade kit yellow | XT-XWHM-Y-NA-K | N/A |
Max XT II upgrade kit Black | XT-XWHM-B-NA-K | N/A |