RKI’s paper tape monitors utilize highly sensitive colorimetric tapes to achieve interference free gas detection. The FP-330 is equipped with an easily replaceable tape that is highly specific to Formaldehyde. This model is equipped with a digital display showing the type of gas and concentration in PPM. The FP-330 will also display remaining tape life as well as an end of tape alarm.

The FP-330 has audible and visual alarms, as well as alarm relay contacts and 4-20mA output. A unique “Tape Saver mode" extends tape life in applications where high background gas levels may be present.

The FP-330 meets and exceeds OSHA and NIOSH compliance requirements, and does not respond to interfering gases (hydrocarbons, alcohols, solvents, CO, etc.) that may be present in industrial/manufacturing environments.

Features:

  • Small and compact
  • Highly sensitive and ideal for low level formaldehyde concentrations
  • Interference-free detection
  • Easy detection tape replacement with plug-in cassette
  • Tape remainder display and tape end warning alarm provided
  • PPM digital display
  • Easily transportable
  • 115 VAC powered
  • Alarm relays
  • 4-20mA output

Applications:

  • Furniture and woodworking
  • Coating and varnishing
  • Engineered wood products
  • Medical / veterinary
  • Painting & corrosion resistant finishes
  • Carpet manufacture & storage
  • Adhesives
  • Heat treatment operations
  • Manufactured homes/trailer/RV inspection

*Please note we can only ship RKI products to customers within the United States

Model FP-330 HCHO 0 - 0.500 ppm FP-330 HCHO 0 - 1.00 ppm FP-330 HCHO 0 - 5.00 ppm
Detection principle
Photoelectric photometry method using tape cassette
Detection tape
Plug in cassette, 65-FCL-018
Detection gas
Formaldehyde (HCHO)
Detection range
0.030 to 0.500 ppm
0.05 to 1.00 ppm
0.3 to 5.00 ppm
Alarm point
1st 0.080 ppm, 2nd 0.300 ppm
1st 0.17 ppm, 2nd, 0.33 ppm
1st 0.8 ppm, 2nd 1.7 ppm
Alarm operation
Latching alarm (standard) or non-latching (option)
Lamp : Caution = Yellow, Alarm = Red, with buzzer
Zero suppression
0.030 ppm
0.05 ppm
0.3 ppm
Measuring cycle
1800 seconds (30 minutes)
600 seconds (10 minutes)
180 seconds (3 minutes)
Resolution
0.005ppm
0.01 ppm
0.1 ppm
Flow rate
450ml/min. (nominal)
Max. tubing length
2m (teflon only)
Accuracy
±20% of indication value + 1 digit
Tape operation time
1 month with no background gas
Alarm Indication
Warning and alarm with audible and visual indications
Self-diagnosis
Low flow, tape cut, sensor failure, etc.
External output signal
4-20mA (load resistance: below 300 Ω)
Operating temperature & humidity
5~35°C 20~60% RH (non condensing)
Gas inlet/outlet
OD 6mm with fittings
Power supply & power
100VAC to 220VAC universal power, ±10%
Dimensions & weight
Approx. 6.5"(W) x 7.8"(H) x 10.4"(D) Approx. 12.125lb
164 (W) x 198 (H) x 263 (D) mm, 6.5kg approx.