1-866-4 JJSTECH(1-866-455-7832)

RKI Instruments FP-260AGZSL Paper Tape Machine, C4F6, Low Range - FP-260AGZSL

Item #:
FP-260AGZSL
Brand:
RKI Instruments
$9,030.00
Qty
Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
RKI Instruments FP-260AGZSL Paper Tape Machine, C4F6, Low Range - FP-260AGZSL

RKI Instruments FP-260AGZSL Paper Tape Machine, C4F6, Low Range - FP-260AGZSL

$9,030.00

FP-260AGZSL Paper Tape Machine, C4F6, Low Range by RKI Instruments

This model is a highly sensitive C5F8 / C4F6 / or C3F6 gas monitor which uses a special techniquie for detecting C5F8, C4F6 or C3F6. into more readily detectable compounds.

Then, the resulting compounds cause a chemical reaction on a colorimetric tape, and the intensity of the chemical reaction is analyzed and correlated to a specific ppm level of C5F8, C4F6 or C3F6 present in the sample. The detection tape is in a handy cassette which is easily removed for periodic replacement.

Features:

  • Ideal for monitoring semiconductor fab clean rooms
  • Ideal for low gas concentration
  • Minimal interference from solvents such as Fluorinert and Galden
  • Easily replaceable plug-in tape cassette
  • Cassette tape remainder display included

Applications:

  • Environmental Monitoring
  • Semiconductor Fabs
  • Manufacturing
  • Detecting uncaptured gas from scrubber

*Please note we can only ship RKI products to customers within the United States

FP-260 Specifications:
Type Desktop type, "Transportable"
Detection principle Special pyrolyzer + photoelectric photometry method
Detection tape Plug-in cassette type
Measuring gas and range C5F8: 0 ~ 5.0ppm
C4F6: 0 ~ 5.0ppm
Detection cycle 20 seconds
Operating time for cassette tape 1 month (for gas free condition)
Design includes tape reminder display and caution display for end of tape.
Sampling method Automatic sampling (1 l/min) with internal pump
Warm up time Approx. 20 minutes
Gas Concentration readout Direct reading with digital display
Alarm Preset Levels C5F8: Warning: 2.0ppm, Alarm: 4.0ppm
C4F6: Warning: 2.0ppm, Alarm: 4.0ppm
Alarm Accuracy Better than ± 30% +1 digit of alarm setting value at constant condition
Alarm Display Warning: Blinking yellow light
Alarm: Blinking red light
(steady light after restart)
Relay Contact Latched mode (standard) or non-latched mode (optional). Form C (dry contacts) for both warning and alarm.
Rated 0.5A at 125VAC.
Self diagnosis Sensor failure, low flow, tape failure, tape cut, tape replacement, tape end, tape failure, tape set failure, system trouble, motor trouble. Output: Form C, dry contacts
non-latched mode, rated 0.5A at 125VAC
Output Signal * 4-20mA DC (0-1V DC optional)
* BL-NET (optional)
Operating temperature & humidity 10 ~ 35°C, 30 ~ 60% RH (non condensing)
Gas Inlet Outer diameter: 6mm, with fitting for PTFE tubing
Power Requirement 100VAC, 110VAC, 115VAC, 220VAC (80VA)
Outer Dimensions Approx. 275(W) x 220(H) x 370(D)mm
Weight Approx. 12 kg
Standard Accessories Gas detection cassette tape, Dust filter, Fuse (3A)

Detectable gases and detection range
Detection gas/chemical formula Range Resolution Warning Alarm Detection Cycle Cassette tape
Octafluorocyclopentene (C5F8) 0~5.0ppm 0.1ppm 2.0ppm 4.0ppm 20 sec FV-017
Perfluorobutadiene (C4F6) 0~5.0ppm 0.1ppm 2.0ppm 4.0ppm 20 sec FV-017
Hexafluoropropylene (C3F6) 0~3.0ppm 0.1ppm 0.9ppm 2.0ppm 20 sec FV-117

Datasheets & User Manuals

Questions & Answers