Features

  • Patented built-in non-contact IR Thermometer design with laser pointer for locating hot spots
  • True RMS CAT III-1000V, CAT IV-600V rated with 0.06% basic DCV accuracy
  • Diode open circuit voltage of 2.8V DC
  • Heavy duty Double molded, water resistant package
  • Large backlit 40,000 count LCD with 40-segment bargraph
  • Memory for storage and recall of three measurements
  • 1000V input protection on all functions
  • Type K thermocouple contact temperature measurements
  • Features include HOLD, RELATIVE and PEAK HOLD
  • Min, Max and Average recording
  • Auto power off with disable feature
  • Capacitance, Frequency, Diode and Continuity functions
  • Complete with double molded test leads, magnetic hanging strap, general purpose Type K bead wire temperature probe, case, and 9V battery
General Specifications
Enclosure Double molded, Waterproof (IP64)
Diode Test Test current of 0.9mA maximum, open circuit voltage 2.8V DC typical
Continuity Check Audible signal will sound if the resistance is less than 35Ω (approx.), test current <0.35mA
PEAK Captures peaks >1ms
Memory Store and recall up to 3 values
Temperature Sensor Requires type K thermocouple
Input Impedance >10MΩ VDC & >3MΩ VAC
AC Response True rms
ACV Bandwidth 40Hz to 1000Hz
IR Spectral response 6 to 16µm
IR Emissivity 0.95 fixed
IR distance ratio 30:1
Laser pointer Class 2 laser < 1mW power; Wavelength is 630 to 670nm
Crest Factor ≤3 at full scale up to 500V, decreasing linearly to ≤1.5 at 1000V
Display 40,000 count backlit liquid crystal display with bargraph
Overrange indication "OL" is displayed
Auto Power Off 15 minutes (approximately) with disable feature
Polarity Automatic (no indication for positive); Minus (-) sign for negative
Measurement Rate 2 times per second, nominal
Low Battery Indication Battery symbol is displayed if battery voltage drops below operating voltage
Battery One 9 volt (NEDA 1604) battery
Fuses mA, µA ranges; 0.5A/1000V ceramic fast blow
A range; 10A/1000V ceramic fast blow
Operating Temperature 5°C to 40°C (41°F to 104°F)
Storage Temperature -20°C to 60°C (-4°F to 140°F)
Operating Humidity Max 80% up to 31°C (87°F) decreasing linearly to 50% at 40°C (104°F)
Storage Humidity <80%
Operating Altitude 2000 meters (7000 ft) maximum
Weight 342 g (0.753 lb) (includes holster)
Size 187 x 81 x 50 mm (7.36" x 3.2" x 2.0") (includes holster)
Safety This meter is intended for origin of installation use and protected, against the users, by double insulation per EN61010-1 and IEC61010-1 2nd Edition (2001) to Category IV 600V and Category III 1000V; Pollution Degree 2.
Approvals CE

Range Specifications
Function
Range
Resolution
Accuracy
DC Voltage
400 mV
0.01 mV
±(0.06% reading + 2 digits)
4V
0.0001V
40V
0.001V
400V
0.01V
1000V
0.1V
AC Voltage
40 Hz to 65 Hz
66 Hz to 1000 Hz
400 mV
0.01 mV
±(0.5% rdg + 3 digits)
±(1% rdg + 3 digits)
4 V
0.0001V
40 V
0.001V
400 V
0.01V
1000 V
0.1V
All AC voltage ranges are specified from 5% of range to 100% of range
DC Current
400μA
0.01 μA
±(0.7% reading + 3 digits)
4000μA
0.1 μA
40mA
0.001 mA
400mA
0.01 mA
±(1.0% reading + 3 digits)
10A
0.00 A
±(1.5% reading + 3 digits)
(20 A: 30 sec max with reduced accuracy)
AC Current
40 Hz to 65 Hz
66 Hz to 1000 Hz
400μA
0.01 μA
±(1.0% rdg + 3 digits)
±(1.5% rdg + 3 digits)
4000μA
0.1 μA
40mA
0.001 mA
400mA
0.01 mA
10A
0.00 A
±(2.0% reading + 3 digits)
±(2.0% reading + 3 digits)
(20A: 30 sec max with reduced accuracy)
Resistance
400Ω
0.01Ω
±(0.3% reading + 4 digits)
4 kΩ
0.0001 kΩ
40 kΩ
0.001 kΩ
400 kΩ
0.01 kΩ
4 MΩ
0.0001 MΩ
40 MΩ
0.001 MΩ
±(2.0% reading + 20 digits)
Capacitance
40 nF
0.001 nF
±(3.5% reading + 40 digits)
400 nF
0.01 nF
4 µF
0.0001 µF
40 µF
0.001 µF
400 µF
0.01 µF
4000 µF
0.1 µF
±(5% reading + 10 digits)
40000 µF
1 µF
Frequency
40 Hz
0.001 Hz
±(0.1% reading + 1 digits)
400 Hz
0.01 Hz
4 kHz
0.0001 kHz
40 kHz
0.001 kHz
400 kHz
0.01 kHz
4 MHz
0.0001 MHz
40 MHz
0.001 MHz
Duty Cycle
0.1 to 99.90%
0.01%
±(1.0% reading + 2 digits)
Pulse width: 100 µs – 100 ms, Frequency: 5 Hz to 150 kHz
Temperature (type-K)
-50 to 1382°F
1°F
±(1.0% reading + 4.5°F)
±(1.0% reading + 2.5°C)
(probe accuracy not included)
-45 to 750°C
1°C
Temperature IR
-30 to -5°C
1°C/°F
±4°C
-6 to -100°C
± 2°C
101 to 550°C
±(2.0%rdg + 2°C)
-22 to 23°F
± 8°F
24 to 212°F
± 4°F
213 to 1022°F
±(2.0%rdg + 4°F)