Casella Microtherm Heat Stress WBGT Kit with Standard Accessories and Calibration Certificate

Item #:
180042A
Brand:
Casella
Casella Microtherm Heat Stress WBGT Kit with Standard Accessories and Calibration Certificate
$2,750.00
Qty

Discontinued

Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Casella Microtherm Heat Stress WBGT Kit with Standard Accessories and Calibration Certificate

Casella Microtherm Heat Stress WBGT Kit with Standard Accessories and Calibration Certificate

$2,750.00

Discontinued


Casella Microtherm Heat Stress WBGT

Casella Microtherm Heat Stress WBGT is the world’s first heat stress meter that offers real time graphical display of data. With its audible and visual alarms, the unit can immediately notify those nearby of pending hazards. The heat stress WBGT has a unique and efficient design that is rugged, compact, and portable. When heat stress dangers are abound, consider Casella and consider the best in heat stress monitoring.

The Dangers of Heat Stress

When a person’s core body temperature rises to a dangerous or hazardous level, the person is under the duress of heat stress. Certain working conditions place workers at risk of this danger by exposing them to hot working environments for prolonged periods of time. Symptoms include heat cramps, nausea, palpitations, stroke, and even death. Using the Wet Bulb Globe Temperature Index (WBGT), workplace managers can provide an overall estimation of heat stress levels on the body. The dangers are very real, and avoidable with the proper tools. The Casella Microtherm Heat Stress WBGT provides a mechanism to monitor potential heat stress.

Key Features of the Heat Stress WBGT

The Microtherm Heat Stress WBGT has several key features that improve its performance in the field. Its large graphic LCD display is very easy to read and provides real time readings. It has a standard feature of full data-logging capabilities (PC software is included), and provides a warning or alarm for low water level and WBGT levels. It has an incredibly efficient ergonomic design and a very long battery life. The unit meets the requirements of ISO 7243. It is also suitable for a variety of applications including

  • Power stations
  • Foundries
  • Steel works
  • Bakeries
  • Glass manufacturing
  • Routine monitoring
  • Medical surveillance testing

Casella Microtherm Heat Stress WBGT Specifications

The WGBT meter simultaneously measures from the wet bulb, dry bulb, and globe temperatures sensors. These sensors are manufactured to the highest accuracy standards. The display indicators are coded for simplicity and require minimal training for use. The data logger has 512k of memory allowing for up to 49,100 data records. The handheld portable device operates on a variety of PC operating systems.

Casella CEL and IDEAL Industries, Inc.

Casella CEL is a wholly owned subsidiary of IDEAL Industries, Inc. Casella has a long history of achievements in the advanced scientific fields dating back to its founding in 1799. With notable clients like Charles Darwin, the company made a name for itself by developing and manufacturing leading scientific apparatus. As science continued to advance, so did Casella. Today, Casella operates under another global leader in science and innovation. IDEAL Industries leads the world in electrical connectors, test equipment, hand tools, and advanced instrumentation.

Identity Sensor specifications
Parameter RANGE ACCURACY
Ta Air temperature 10-60˚C +/- 1˚C
Tg Globe temperature 20-120˚C +/- 0.5˚C (20-50)
+/-1.0˚C (50-120)
Tnw Natural wet temperature 5 to 40˚C +/- 0.5˚C
Transducers PRTD 100 elements 0.1% using 4 wire extension cables available in 10m, lengths up to 30m
Displayed Data Values
Ta Air temperature
Tg Globe temperature
Tnw Natural wet temperature
WBGT Inside and Outside
Time Weighted WBGT Values Based on a 1 hr rolling average updated every 30 seconds. (During the first hour ‘— -’ is displayed.)
%RH Based on”Ta” and non aspirated “Tnw” sensor
Dewpoint Based on “Ta” and non aspirated “Tnw” sensor
Data Logger
Total Record Capacity 512k memory providing up to 49,100 data records
Number of Runs Stored 32
Logging Interval 30 seconds to 1 hour
Software
Operating System MS Windows 95/98 NT, XP, Vista compatible. RS232 serial port, 8MB RAM, FDD & HDD, VGA Graphics.
Power
Battery Power supply 4 x AA cells Dry cells or NiMH
Battery life 40hrs NiMH / 85hrs alkaline. Battery monitoring warning and auto power down. Consumption; 30mA typical
External power supply 3.5 to 14VDC (12v nominal) 90-240VAC Universal input mains power adaptor. Internal fast charge circuitry
Communications
RS232  
Applicable Standards
ISO 7243/7726, OSHA  
Operating Temperature Range
Sensors -5˚C to +120˚C
Electronics -5˚to +60˚C
Dimensions Instrument: 245 x 95 x 50mm. Array: 90 x 225 x 65mm. Sensor/carry case (H x W xD): 135 x 490 x 370mm
Weight Sensor/carry case 0.97kg. Instrument only 0.75kg