- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Sensor Configuration
Sensor Configuration
The Sensor Configuration tab adjusts settings for each individual sensor. A separate sensor tab is provided for each sensor. Figure 5. shows the available option settings for the H2S sensor.
Figure 5. Sensor Configuration Tab (H2S)
Depending upon the sensor, the options may vary
Sensor Disabled
Warning
Use extreme caution when disabling a sensor. The disabled sensor cannot detect and alarm against the applicable gas.
To disable a sensor, complete the following:
1. Click Retrieve from Device to populate the fields with the current detector settings.
2. Click the tab of the sensor to be disabled.
3. Click the checkbox for Sensor Disabled.
When options are enabled/disabled, the checkbox displays with a red frame until the new settings are saved to the detector.
4. Click the Save to Device button located at the bottom of the window.
5. The LCD automatically updates. In the following example, the CO gas type and sensor readings no longer displays.
Calibration Gas (ppm)
Warning
The gas concentration value entered in Fleet Manager II must match the gas concentration value on the gas cylinder.
1. Select the applicable sensor tab.
2. Enter the gas concentration value in the Calibration Gas (ppm) field for H2S and CO.
3. Enter the gas concentration value in the Calibration Gas (%) field for O2 and LEL.
The gas concentration value for the O2 sensor must be calibrated with a %value other than 20.9 or 20.8%, such as 18%.
Calibration Interval
Define how often a sensor must be calibrated in the Calibration Interval (days) field. A different calibration interval can be set for each sensor.
1. Enter the value (0-365 days) for each sensor.
2. Enter 0 to disable the calibration interval option. Entering 0 automatically deactivates the Force Calibration user option.
BW recommends that the detector be calibrated at least once every 180 days (6 months).
The detector is shipped with the factory default set to 180 days.
Bump Interval
Define how often a bump test must be performed for each sensor in the Bump Interval (days) field. A different bump interval can be set for each sensor.
1. Enter the value (1-365 days) for each sensor.
2. Enter 0 to disable the bump interval option. Entering 0 automatically deactivates the Force Bump user option. The detector is shipped with the Force Interval option disabled.
BW recommends to bump test the sensors before each day