- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Replace the Pump
Replace the Pump
1.Turn the instrument off.
2.Remove the battery:
a) Turn the instrument face down and unscrew on the battery pack.
b) Remove the battery pack and the four screws in the battery compartment
3.Remove the two screws at the top of the instrument and pull the back shell apart from the front shell.
4.Remove the pump:
a) Pull the pump out of the manifold assembly.
b) Release the pump connector retainer clip and pull the pump connector out.
5.Install the new pump:
a) Retrieve the new pump and insert the pump connector.
b) Fold the pump wires across the bottom of the pump motor.
c) Insert the pump into the manifold assembly
6.Replace the back shell and install the 4 screws back into the battery compartment (torque 4-5 in-lbs).
7.Reinstall the two screws at the top of the instrument (torque 4-5 in-lbs).
8.To place the battery pack correctly, first engage hooks on the bottom of the battery pack
9.Push the battery pack into place and tighten the battery retention screw (torque 4-5 in-lbs).