- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Replace the 4R+ Sensors
Replace the 4R+ Sensors
1.Turn the instrument off.
2.Remove the battery:
a) Turn the instrument face down and unscrew on the battery pack.
b) Remove the battery pack and the four screws in the battery compartment
3.Remove the sensor:
a) Remove the two screws at the top of the instrument and pull the back shell apart from the front shell.
b) Disconnect the 4R+ cable from the sensor.
c) Push at the PRESS marking in the manifold toward the top of the instrument.
d) Use the tabs on the sensor to pull it out of the manifold.
4.Install the new sensor:
a) Retrieve the new sensor and insert it into the manifold aligning the sensor guides to the instrument