- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Replace the 1-Series Sensors
Replace the 1-Series Sensors
1.Turn the instrument off.
2.Remove the battery:

a) Turn the instrument face down and unscrew on the battery pack.

b) Remove the battery pack and the four screws in the battery compartment
3.Remove the sensor:

a) Remove the two screws at the top of the instrument and pull the back shell apart from the front shell.

b) Pull back the two retaining clips on the manifold.

c) Lift the manifold sideways, farthest from the pump.

d) Pull the sensor manifold out from the pump manifold.

e) Remove the desired sensor.
4.Install the new sensor:

a) Mind the orientation key, and then push the sensor down into manifold.

b) Insert manifold inlet into pump manifold.
c) Push manifold down and as the two latches engage, listen for two snapping sounds

5. Replace the back shell and install the four screws back into the battery compartment (torque 4-5 in-lbs).

6. Reinstall the 2 screws at the top of the instrument (torque 4-5 in-lbs)
7.Attach the battery:

a) To place the battery pack correctly, first engage hooks on the bottom of the battery pack.
b) Push the battery pack into place and tighten the battery retention screw (torque 4-5 in-lbs).
8.Turn on the instrument and allow for the sensors to stabilize.