pH/EC 80 pH/Conductivity/Temperature Meter by Jenco

Multi-parameter tester for measuring pH, conductivity, TDS, salinity and temperature

The pH/EC80 is one of the most advanced multi-parameter tester on the market. The tester has a large graphic LCD that simultaneously displays pH or EC measurement and temperature measurement. Users can store up to 50 memory data sets for recall and select degrees Celsius or Fahrenheit as the temperature unit of measure. In addition, the advanced hardware and the user-replaceable electrode module are protected by an IP-67 rated housing.

  • Simultaneously measures pH, Conductivity, TDS, Salinity and temperature without switching probe
  • Graphic LCD display
  • Auto temperature compensation
  • Switches between °C and °F
  • Auto shut down feature
  • Replaceable pH/Conductivity electrode module
  • 50 sets of data memory
  • IP67 waterproof
  • Battery life >100hrs
  • CE approved and RoHS compliant
pH Range 0.00 to 14.00
pH Resolution 0.01
pH Accuracy ± 0.01 pH
Conductivity Range Range1 0 uS/cm to1999 uS/cm
Conductivity Resolution Range1 1 uS/cm
Conductivity Accuracy Range1 ±1% F.S.
Conductivity Range 2 2.00 mS/cm to 20.00 mS/cm
Conductivity Resolution 2 0.01 mS/cm
Conductivity Accuracy 2 ±1% F.S.
Salinity Range 0.0 to 10.0 ppt
Salinity Resolution 0.1 ppt
Salinity Accuracy ±1% F.S.
TDS Range 1 0 to 1999 mg/L
TDS Resolution 1 1 mg/L
TDS Accuracy 1 ±1% F.S.
TDS Range 2 2.00 to 20.00 g/L
TDS Resolution 2 0.01 g/L
TDS Accuracy 2 ±1% F.S.
Temperature Range 1 -5.0 to 60.0 °C
Temperature Resolution 1 0.1 °C
Temperature Accuracy 1 ±0.5 °C
Temperature Range 2 23 to 140 °F
Temperature Resolution 2 0.1 °F
Temperature Accuracy 2 ±0.9 °F
pH Calibration 1, 2 and 3 point calibration. Uses USA and NIST buffer sets.
Conductivity Calibration 1 point calibration
Temperature Compensation Range -5.0 to 60.0 °C
Referemce Temperature 15 to 25 °C Default at 25 °C
Temperature Coefficient 0.00 to 4.00 %. Default at 1.91%/°C
TDS Constant 0.30 to 1.00, Default at 0.65
Temp Sensor Thermistor, 10k ohms, at 25 °C
Temp Unit °C and °F
Input Impedance >1012 Ohms
Display Graphic LCD
Storage of Calibration Data Yes
Data Memory Storage 50 sets
Auto Shutoff After 10 minutes of inactivity.
Power 4 x LR44 batteries
Case IP67 water-tight case
Working Environment 0 to 50 °C, relative humidity < 95 %
Dimensions 200 × 44 × 25 mm
Weight About 110 g (excluding batteries)