Reconfiguring the Detector (not applicable to the GasAlertClip Extreme)
Fleet Manager II can reconfigure the following detectors with the base station:
- GasAlert Extreme
- GasAlertMicro
- GasAlertMicro 5/PID/IR
- GasAlertMicroClip
- GasAlertMax XT
The detector can be reconfigured prior to performing a bump test or calibration.
To reconfigure a detector, complete the following:
- Activate the base station. Ensure it is in normal operation
- Ensure that a correctly formatted MMC/SD card is inserted in the base station. Refer to Inserting/Replacing a MMC/SD Card.
- Connect the USB cable to the computer and to the USB port on the base station.
- Activate the detector and wait until it is in normal operation
- Insert the detector into the detector bay.
- From the computer, open Fleet Manager II.
- Login to the Administration functions
- Under the Devices tab, click Configure Devices.
- The Load Configuration File dialog box displays.
Select one of the following options:
- Load existing configuration file from a MicroDock
- Load existing configuration file from hard drive
- Load default configuration file
- The Device Configuration dialog box displays.
- Select which device to configure by clicking its tab.
- Configure the device
Ensure the Updatable checkbox is checked.
- Select a program option that will reprogram the detector. Each detector