Order of Gases Applied for Bump Tests and
Calibrations
Warning
It is extremely important that gases are applied in the order that is specified in this manual.
Failure to adhere to the rules in Table 18. and Table 19. of this manual will result in incorrect sensor readings and possible personal injury and/or property damage.
Note
Manual application of gas types is available for the GasAlertMicro 5/PID/IR detectors only
To manually apply gases for bump tests and calibrations, complete the following:
- Review Table 18. and Table 19.
- Determine which gases will be applied.
- Using Table 18., locate the first gas type (sensor) to be bump tested or calibrated on the top row.
- Within the far left column, locate the next gas type (sensor) to be bumped or calibrated.
The square where the column and row intersect will indicate which gas should be applied first and a rule number (if applicable).
- Using Table 19., locate the corresponding rule number. Ensure that the desired order to apply gases corresponds to the gas application rules.
- Continue to reference Table 18. and Table 19. to ensure the gases will be applied in the correct order.
- When the correct order is determined, proceed to the Bump Test or Calibration section.
Table 18. Gas Type Application Table
Gases |
H2S |
CO |
CI2 (Bump only |
NH3 |
HCN |
NO2 |
COSH (H2S/CO) |
SO2 |
PH3 |
PID |
H2S |
|
Quad gas |
CI2 Rule 4 |
NH3 |
HCN |
NO2 |
quad gas - one step calibration |
SO2 |
PH3 Rule 11 |
H2S |
CO |
|
|
CI2 Rule 5 |
NH3 |
HCN |
NO2 |
quad gas - one step calibration |
SO2 |
PH3 Rule 11 |
CO |
CI2 Bump only |
|
|
|
NH3 Rule 9 |
HCN |
NO2 Rules 8 &10 |
CI2 Rule 6 |
CI2 Rule 9 |
PH3Rule 9 |
CI2 |
NH3 |
|
|
|
|
HCNRule 9 |
NO2 |
NH3Rule 6 |
NH3Rule 9 |
PH3Rules 9 &10 |
NH3 |
HCN |
|
|
|
|
|
HCN |
HCNRule 6 |
HCNRules 8 &10 |
PH3Rules 8 &10 |
HCN |
NO2 |
|
|
|
|
|
|
NO2 Rule 6 |
NO2 |
PNO2 Rule 1 |
NO2 |
COSHH2S/CO |
|
|
|
|
|
|
|
SO2Rule 6 |
PH3Rules 10 &11 |
COSHH2S/CO |
SO2 |
|
|
|
|
|
|
|
|
PH3Rules 9 &10 |
SO2 |
PH3 |
|
|
|
|
|
|
|
|
|
PH3 |
PID |
|
|
|
|
|
|
|
|
|
|
Table 19. Gas Application Rules
Rule # |
Gas Type |
Apply |
Exception(s) |
1 |
PH3 |
First |
Two exceptions: NO2 and O3 |
2 |
HCN |
First |
Three exceptions: PH3 and ClO2 |
3 |
NO2 |
First |
Three exceptions: HCN |
4 |
H2S |
Last |
One exception: PID |
5 |
CO |
Last |
One exception: PID |
6 |
COSH (H2S/CO) |
Last |
One exception: PID |
7 |
PID |
Last |
No exceptions |
8 |
Toxic gas |
|
Both toxic sensors are cross sensitive to each other
|