Order of Gases Applied for Bump Tests and Calibrations


Order of Gases Applied for Bump Tests and Calibrations

Warning

It is extremely important that gases are applied in the order that is specified in this manual.
Failure to adhere to the rules in Table 18. and Table 19. of this manual will result in incorrect sensor readings and possible personal injury and/or property damage.

Note

Manual application of gas types is available for the GasAlertMicro 5/PID/IR detectors only

To manually apply gases for bump tests and calibrations, complete the following:

  1. Review Table 18. and Table 19.
  2. Determine which gases will be applied.
  3. Using Table 18., locate the first gas type (sensor) to be bump tested or calibrated on the top row.
  4. Within the far left column, locate the next gas type (sensor) to be bumped or calibrated.
  5. The square where the column and row intersect will indicate which gas should be applied first and a rule number (if applicable).

  6. Using Table 19., locate the corresponding rule number. Ensure that the desired order to apply gases corresponds to the gas application rules.
  7. Continue to reference Table 18. and Table 19. to ensure the gases will be applied in the correct order.
  8. When the correct order is determined, proceed to the Bump Test or Calibration section.

Table 18. Gas Type Application Table

Gases H2S CO CI2 (Bump only NH3 HCN NO2 COSH (H2S/CO) SO2 PH3 PID
H2S Quad gas CI2 Rule 4 NH3 HCN NO2 quad gas - one step calibration SO2 PH3 Rule 11 H2S
CO CI2 Rule 5 NH3 HCN NO2 quad gas - one step calibration SO2 PH3 Rule 11 CO
CI2 Bump only NH3 Rule 9 HCN NO2 Rules 8 &10 CI2 Rule 6 CI2 Rule 9 PH3Rule 9 CI2
NH3 HCNRule 9 NO2 NH3Rule 6 NH3Rule 9 PH3Rules 9 &10 NH3
HCN HCN HCNRule 6 HCNRules 8 &10 PH3Rules 8 &10 HCN
NO2 NO2 Rule 6 NO2 PNO2 Rule 1 NO2
COSHH2S/CO SO2Rule 6 PH3Rules 10 &11 COSHH2S/CO
SO2 PH3Rules 9 &10 SO2
PH3 PH3
PID

Table 19. Gas Application Rules

Rule # Gas Type Apply Exception(s)
1 PH3 First Two exceptions: NO2 and O3
2 HCN First Three exceptions: PH3 and ClO2
3 NO2 First Three exceptions: HCN
4 H2S Last One exception: PID
5 CO Last One exception: PID
6 COSH (H2S/CO) Last One exception: PID
7 PID Last No exceptions
8 Toxic gas Both toxic sensors are cross sensitive to each other