MicroDock II Advanced Procedures

MicroDock II Factory mode

  • Factory Mode used to update inlet configurations after a firmware update
  • To enter factory mode
    • Press and hold two center buttons on main module
    • Then press left most button and hold all three
MicroDock II Factory mode

Module Initialization

  • Primary reason for module initialization is new module installation
    • Initialization has been proven to be a fix for many docking station problems
  • To initialize:
    • Power down docking station
    • Attach module
      • Ensure both gas and electrical connections are made
    • Press and hold bump check button on newly added module
      • Then press and hold left most or power button on Main Module
    • After a few seconds message "New receptacle added at position __" will appear
    • Closest module to main is 1 and so on

Re-Initialization

  • If a dock is experiencing problems
    • Check firmware revisions in user options for strange characters or missing modules
      • Inside firmware user options screen there should be one line for main module and one for each dock module
  • Initialization has one limitation, only distal module can be initialized
  • If you are working on a dock with 5 instrument modules and the first module is having problems you will have to remove the other 4 initialize the first, attach the second initialize and so forth.