Dimensions
14.2 x 5.4 x 4.3 in (361 x 137 x 109 mm)
Weight
4.2 lb (1.9 kg)
Base system ingress protection
IP20
Enclosure
Impact resistant polycarbonate and acrylonitrile butadiene styrene (PC+ABS)
Operating temperature
+50°F to +95°F (+10°C to +35°C)
Humidity
10% to 90% non-condensing
Altitude
Maximum 6 561 feet (2 000 meters)
Storage temperature
+14°F to +140°F (-10°C to +60°C)
Power supply
12 V 8.33A ITE power supply
Type of equipment
Process Control Equipment
Equipment class
Cord-connected, Class III
Pollution degree
2
Installation category
I
Mode of operation
Continuous
Real-time clock
Provides date and time stamp
Data storage
2 gigabytes of on-board memory
External interface
Ethernet
Pump flow rate
350 mL per minute
Calibration gas cylinder inputs
1 or 4 gas inlets
1 ambient air (purge) inlet
Automatic tests
Functional bump test; calibration; data transfer; audible alarm; visual alarm
Configuration recognition
Automatic instrument and sensor
Alarm/calibration parameters
User-defined
Calibration gas connections
Built-in
Gas connection
3.2 mm (1/8 in) sub-miniature connect (SMC) coupling
Solenoid
Built-in
LCD Background
Gray: Idle
Blue: Prompt for user action
Yellow: Activity in progress
Green: Activity completed
Red: Warning
Keypad
Scroll right
Scroll left
Scroll up
Scroll down
Press and hold until the module settings menu is displayed. Press and release to select a menu item or save changes.
Press and hold until an operation is cancelled. Press and release to close a menu item or return to a previous screen or cancel changes.
Communications method
Two-way infrared between docking module and detector
Ports
1 USB port
1 Ethernet port, for connection to a computer network
Sensors
Audio and optical
Warranty
2 years beginning at date of shipment to the buyer
FCC
NOTE: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
FCC CAUTION: Any changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.