General Repair Guidelines
- Prior to returning a monitor to customer always
- Calibrate
- Bump and check for stability
- By rules of intrinsic safety
- Sensor Warranties
- NH3, ETO, ClO2, Cl2,O3, sensors carry 1 year warranty
- 3 year Clip Extremes have a 3 year warranty
- All others are 2 year
- Corroded monitors are not covered by warranty
- Not to be repaired
- Owner must buy new replacement
Universal Repair Tips
- If monitor display is missing segments
- Replace sensor if
- Sensor shows fail
- Fails to bump and calibrate
- Values constantly float up and down
- If unit has erratic toxic sensor
- Try airing out over a weekend and then recalibrating
- Often fixes poisoned sensors
- When reassembling a monitor
- Always turn counter clockwise until a click is felt then tighten
- If monitor has more than three stripped screws replace case
- 180 days is default calibration interval for GasAlert Family
Factors Affecting Battery Life
- Amount of time unit is in alarm
- How often backlight is used
- Operating Temperature
- Colder temperatures significantly shorten battery life
- Age of rechargeable battery
- Initial charge/discharge cycle
- Rechargeable batteries should be fully discharged then fully charged again 3 times to ensure maximum life
Monitor Cleaning
- Approved cleaners for BW monitors
- ACL Staticide
- Warm water
- Do not use any other cleaners such as Econoclean citrusbased cleaners or Armor All
- To clean monitor
- Apply cleaner of choice to clean cloth and wipe unit clean
- Avoid exposing sensors screens to cleaning liquid