Instrument dimensions: 12.9 x 8.0 x 3.8 cm (5.1 x 3.2 x 1.5 in.)
Weight: 317 g (11.2 oz.)
Operating temperature: -20°C to +50°C (-4°F to +122°F)
Storage temperature: -40°C to +60°C (-40°F to +140°F)
Operating humidity: 10% to 100% relative humidity(non-condensing)
Dust and moisture ingress: IP66/67 (with screw engaged)
Alarm setpoints: May vary by region and are user-defined

Detection range:
H2S: 0 - 200 ppm (1 / 0.1 ppm increments)
CO: 0 - 1000 ppm (1 ppm increments)
O2: 0 - 30.0% vol. (0.1% vol. increments)
Combustible (LEL): 0 - 100% (1% LEL increments) or 0 - 5.0% v/v methane

Sensor type: H2S, CO, O2: Single plug-in electrochemical cell Combustibles: Plug-in catalytic bead
O2 measuring principle: Capillary controlled concentration sensor
Bump test specified limits: BW recommends using a gas cylinder that will ensure the combustible sensor has an accuracy of -0 to +20% of actual reading
Alarm conditions: TWA alarm, STEL alarm, low alarm, high alarm, multi alarm, over limit (OL) alarm, low battery alarm, critical low battery alarm, confidence flash, confidence beep
Audible alarm: 95 dB at 30 cm (12 in.) variable pulsed beeper ber after the second letter determines the year of manufacture. E.g., QA109-001000 = 2009 year of manufacture
Approved lithium battery for GasAlertQuattro product:Lithium-ion polymer (QT-BAT-R01) as per standards UL913, EN60079-11, EN60079-0, C22.2 No. 157
Rechargeable battery (QT-BAT-R01)     Temperature code
Lithium polymer -20°C ≤ Ta ≤ +50°C      T4

Lithium battery operating time: One rechargeable lithium polymer battery provides the following operating runtimes:

20 hours at 20°C (68°F)
18 hours at -20°C (-4°F)

Approved alkaline battery pack for GasAlertQuattro (QT-BAT-A01): as per standards UL913, EN60079-11, EN60079-0, C22.2 No. 157
Approved alkaline batteries for GasAlertQuattro product:
Duracell MN1500        -20°C ≤ Ta ≤ +50°C        T4 (129.9°C)
Energizer E91VP        -20°C ≤ Ta ≤ +50°C        T3C (135.3°C)

AA Alkaline battery operating time: 14 hours at 20°C (68°F)
Battery charger:Charging adapter
First-time charge: 6 hours
Normal charge: 6 hours
Warranty:2 years including sensors

Visual alarm: Red light-emitting diodes (LEDs)
Confidence flash: Green light-emitting diode. Flash frequency is user-defined with confidence flash interval option
Confidence beep: Audible beep from variable pulsed beeper. Beep frequency is user-defined with confidence beep interval option

Display: Alphanumeric liquid crystal display (LCD) with flip display (0° or 180°) capability (user-defined in Fleet Manager II)
Backlight:Activates upon startup and deactivates when selftest is complete. Activates when the pushbutton is pressed and deactivates after 10 seconds. Also activates during an alarm condition and remains lit until alarm ceases

Internal vibrator:Vibrates during activation, deactivation, and all alarms
Self-test: Initiated during activation, self-test runs continuously while detector is operational
Calibration: Zero and automatic span
User options:Startup message, lockout on self-test error, safe mode, confidence flash, confidence beep, latching alarms, force calibration, force bump, calibration IR lock, flip display stealth mode, datalog interval, confidence flash interval, confidence beep interval, and language selection
Sensor options: Sensor enable/disable, calibration gas values, calibration interval, bump test interval, alarm setpoints (low/high/TWA/STEL), STEL interval, TWA period, auto zero at startup enable/disable, LEL correction factor, 10% (of reading) over-span, low alarm acknowledge, O2 measurement, LEL gas measurement, %vol methane measurement

Year of manufacture: The detector's year of manufacture is determined from the serial number. The second and third num-

Approvals:Approved by CSA to both U.S. and Canadian Standards CAN/CSA C22.2 No. 157 and C22.2 152
ANS/UL - 913 and ANSI/ISA - S12.13 Part 1

CSA Class I, Division 1, Group A, B, C, D
ATEX CE 0539 II 1 G Ga Ex ia IIC T4 for Zone 0
Group IIC
KEMA 09 ATEX 0137
EN 60079-0, EN 60079-11, EN 60079-26
IECEx Ga Ex ia IIC T4 IECEx CSA 09.0006
IEC 60079-0, IEC 60079-11, IEC 60079-26

This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 Canadian EMI requirements. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the receiving antenna.
  • Increase the separation between the equipment and receiver
  • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
  • Consult the dealer or an experienced radio/TV technician for help.