Troubleshooting
If a problem occurs, refer to the solutions provided in Table 21.
If the problem persists, contact JJS Technical Services.
Table 21. Troubleshooting Tips
Table 21. Troubleshooting Tips (cont.)
Problem |
Possible Cause |
Solution |
|
The MMC/SD is not inserted |
Insert the MMC/SD card. Refer to Inserting the MMC/SD Card. |
|
The MMC/SD card that is inserted in the detector has a storage size that is not supported by the detector |
Insert an Infineon MMC/SD card that is |
|
The detector has lost communication with the MMC/SD card |
Attempt the following solutions:- Retry communication
- Insert a different Infineon MMC 32 MB
- Clear MMC in windows and reinsert
into the detector.
- Contact JJS Technical Services.
|
Table 21. Troubleshooting Tips (cont.)
Problem |
Possible Cause |
Solution |
The detector displays a clock error message using last recorded time. |
General fault |
Reactivate the detector. If the message does not display, reset the clock in user options. If the error message still displays, contact JJS Technical Services. |
Detector does not display normal ambient gas reading after activation self-test. |
Sensor not stabilized
Detector requires calibration
Target gas is present |
Used sensor: wait 60 seconds New sensor: wait 5 minutes Refer to Calibration and Setting Alarm Setpoints. Detector is operating properly. Use caution
in suspect areas. |
Detector does not respond to pushbuttons. |
Batteries are depleted
Detector is performing operations that do not require user input. |
Refer to Replacing/Charging the Batteries.
Pushbutton function restored automatically when the operation ends. |
Detector does not accurately measure gas. |
Detector requires calibration
Detector is colder/hotter than ambient gas
Sensor filter is blocked |
Refer to Calibration and Setting Alarm Setpoints. Allow the detector to acquire ambient temperature before using.
Clean the sensor filter. Refer to Replacing a Sensor or Sensor Filter. |
Table 21. Troubleshooting Tips (cont.)