Fleet Manager II Configuration
- Use the latest Fleet Manager II
- Log in as Admin with the password Admin
- Select Devices and Configure VIA IR LINK
- Requires same drivers as Softools
Fleet Manager II Configuration Basics
Firmware
- To upgrade
- Connect with Fleet Manager II
- Click boot loader
- Installed Firmware version
- Available during start up
- After "retrieve form device" on Fleet Manager II
- Version 8 is latest
- Compatible with both Max XT and Max XT II
LEL Sensor
- Monitor uses Micropel75C sensor
- Concentration can be viewed as percent by volume of methane or as %LEL
- Can be set to over span by 5%
- Instead of setting calibration gas to 55%.
- Correction factor is available
- Multiplier used to calibrate with a surrogate gas to show accurate readings for another gas
Oxygen Sensor
- Uses the 9 series O2 sensor rated to -20c
- Can be set to zero on 20.8% instead of 20.9%
Toxic Sensors
- Same sensors used in GasAlertMicroClip
- All alarms and STEL/TWA timers are settable
- Defaults to auto zeroing on start up
Event Logs
- All events recorded similar to GasAlertMicroClip
- New view button allows to select which columns to view
Data Logs
- Data logs keep track of all basic device operations