Fleet Manager II Configuration

  • Use the latest Fleet Manager II
    • Log in as Admin with the password Admin
    • Select Devices and Configure VIA IR LINK
    • Requires same drivers as Softools
Fleet Manager II Configuration
Fleet Manager II Configuration

Fleet Manager II Configuration Basics

Fleet Manager II Configuration Basics

Firmware

  • To upgrade
    • Connect with Fleet Manager II
    • Click boot loader
      • Find file and send it
  • Installed Firmware version
    • Available during start up
    • After "retrieve form device" on Fleet Manager II
  • Version 8 is latest
    • Compatible with both Max XT and Max XT II
Firmware
Firmware

LEL Sensor

  • Monitor uses Micropel75C sensor
    • Concentration can be viewed as percent by volume of methane or as %LEL
  • Can be set to over span by 5%
    • Instead of setting calibration gas to 55%.
  • Correction factor is available
    • Multiplier used to calibrate with a surrogate gas to show accurate readings for another gas
LEL Sensor

Oxygen Sensor

  • Uses the 9 series O2 sensor rated to -20c
  • Can be set to zero on 20.8% instead of 20.9%
Oxygen Sensor

Toxic Sensors

  • Same sensors used in GasAlertMicroClip
    • All alarms and STEL/TWA timers are settable
  • Defaults to auto zeroing on start up
Firmware
Firmware

Event Logs

  • All events recorded similar to GasAlertMicroClip
    • New view button allows to select which columns to view
Event Logs

Data Logs

  • Data logs keep track of all basic device operations
Data Logs