Problem | Possible Cause | Solution |
Detector does not activate |
No battery |
Depleted battery |
Damaged or defective detector |
Reversed battery |
|
Install battery |
Replace battery |
Contact BW Technologies by Honeywell |
Reinstall the battery correctly |
|
Detector enters alarm mode immediately when it is activated
|
Sensor needs to stabilize |
Low battery alarm |
Sensor alarm |
|
Used sensor: wait 60 seconds. New sensor: wait 5 minutes |
Replace battery |
Replace sensor |
|
Start up self-test fails during checks
|
General fault |
Alarm setpoints incorrect |
|
Contact BW Technologies by Honeywell |
Reset alarm setpoints |
|
Detector does not display a normal ambient gas reading after activation self-test |
Target gas is present |
Detector requires calibration |
Sensor not stabilized |
|
Detector is operating properly. Use caution in suspect areas |
Calibrate detector |
Used sensor: wait 60 seconds New sensor: wait 5 minutes |
|
Detector does not respond to the push-buttons |
Battery is depleted |
Detector is performing operations that does not require user input |
|
Replace battery |
Pushbutton operation restores automatically when operation ends |
|
Detector does not accurately measure gas |
Detector requires calibration |
Detector is colder/hotter than ambient gas |
Sensor screen is blocked |
|
Calibrate sensor |
Allow detector to acquire ambient temperature before use |
Clean sensor screen |
|
Detector does not enter alarm mode |
Alarm setpoint(s) are set incorrectly |
Alarm setpoint(s) set to zero |
Detector is in calibration mode |
|
Reset alarm setpoints |
Reset alarm setpoints |
Complete calibration |
|
Detector intermittently enters alarm mode without apparent reason |
Ambient gas levels are near alarm setpoint or sensor is exposed to a puff of the target gas |
Alarms set incorrectly |
Missing or faulty sensor |
|
Detector is operating normally. Use caution in suspect areas. Check MAX gas exposure reading |
Reset alarm setpoints |
Replace sensor |
|
Detector automatically deactivates |
Automatic shutdown feature activated due to depleted battery | |
|
Detector does not auto zero or calibrate |
|
|
O2 sensor reading is erratic |
|
|