Display Elements

Display Elements

Parts of the GasAlertMax XT II

Figure 2. Display Elements

Table 4. Display Elements

ItemDescription
1Alarm condition
2 Automatically zero sensor indicator
3 Numeric values
4Battery life indicator
5Pump indicator
6 Heartbeat indicator
7Gas type identifiers
8Gas cylinder indicator
9Automatically span sensor indicator

Pushbutton

PushbuttonDescription
Gas alarm
  • To activate the detector press Gas alarm.
  • To deactivate the detector, press and hold Gas alarm until the OFF countdown is complete and the LCD deactivates.
  • To view the date/time, TWA, STEL, and maximum (MAX) readings, press Gas alarm twice rapidly. To clear the TWA, STEL, and MAX readings, press Gas alarm when the LCD displays CLEAR ALL.
  • To initiate calibration, press and hold Gas alarm as the detector performs the OFF countdown. Continue holding Gas alarm while the LCD briefly deactivates and then reactivates to begin the CAL countdown. Release Gas alarm when the CAL countdown is complete.
  • To activate the backlight, press Gas alarm and release.
  • To acknowledge the latched and pump alarms, press Gas alarm.
  • To acknowledge a low alarm and disable the audible alarm temporarily, press Gas alarm. The Low Alarm Acknowledge option must be enabled in Fleet Manager II
  • To acknowledge any of the Due Today alarms (calibration, bump test, pump block test), press Gas alarm.