Calibration

Calibration

CAUTION

Calibrate only in a safe area that is free of hazardous gas in an atmosphere of 20.9 % oxygen. Do not calibrate the detector during or immediately after charging is complete

Note The maximum hose length for calibration is 3 ft. (1 m).

Gas alarm1. Press and hold Gas alarm as the detector performs the OFF countdown. Continue to hold Gas alarm when the detector briefly deactivates.

Gas alarm2. The detector activates again and performs the CAL countdown. Continue to hold Gas alarm until the countdown is complete to enter calibration.

Gas alarm3. flashes while the detector zeroes all of the sensors. If a sensor fails to auto zero, it cannot be spanned. When auto zero is complete, the LCD displays APPLY GAS.

4. Refer to Calibration and Bump Test Installation. Attach the 0.5 l/min. regulator or the demand flow regulator and apply gas

NoteThe diffusion cover must be attached to the detector to calibrate.

Gas alarm flashes on the LCD. After a sufficient amount of gas has been detected (30 seconds), the detector beeps. Auto Span flashes while the detector completes the span.

Gas alarm5. The LCD displays CAL DUE. Next, a screen displays showing the number of days remaining before calibration is due for each sensor. Last, the LCD displays (e.g.) CAL DUE 180 d. As some sensors require more frequent calibrations, the LCD displays the earliest calibration that must be performed.

Note Calibration can be aborted at any time. To abort calibration, press Gas alarm. The CAL ABORTED screen displays.