1-866-4 JJSTECH(1-866-455-7832)

Calibration Principles

Calibration

  • Verify accuracy on a regular basis is to guard against any unexpected loss of sensitivity
  • Document!
Calibration

Mandatory to use a "calibrated" instrument maintained according to "manufacturer requirements"

Calibration
  • 1910.146(c)(5)(ii)(C):
    • Before an employee enters the space, the internal atmosphere shall be tested, with a calibrated directreading instrument
    • What does OSHA accept as a "calibrated" direct reading instrument?
      • A testing instrument maintained and calibrated in accordance with the manufacturer's recommendations
      • The best way for an employer to verify calibration is through documentation

Why do instruments need to be tested and / or calibrated?

  • The response of gas detecting sensors can change over the life of the sensor
  • The changes may be sudden, or can be gradual
  • Substances or conditions present in the atmosphere can have an adverse effect on the sensor
  • Different types of sensors have different constraints and conditions which can lead to loss of sensitivity or failure
  • Important to know how sensors detect gas to understand conditions that can lead to inaccurate readings
instruments need to be tested

Make sure the instrument has been calibrated!

instrument
  • Follow manufacturer recommendations
  • "Zero" instrument in fresh air prior to use
  • Verify Accuracy Daily!
  • Functional "bump" test sufficient
  • Adjust "span" only if necessary

Loss of sensitivity can be due to:

  • Aging or desiccation of the sensors,
  • Mechanical damage due to dropping or immersion
  • Exposure to sensor poisons present in the atmosphere being monitored
  • Loss of sensitivity due to other causest
instrument

Regulatory Requirements

instrument
  • OSHA 1910.146 requires use of a "calibrated" instrument
  • This means (per OSHA CPL 2.100) that the instrument must be maintained and calibrated according to manufacturer guidelines

Manufacturer Guidelines

  • OSHA holds instrument users accountable to maintain, calibrate and operate their instruments according to manufacturer guidelines
Manufacturer Guidelines

UL / CSA Product Markings

  • Typical North American Marking:
    • c-CSA-us or c-UL-us Classified as to intrinsic safety for use in Class I, Division1 Groups A, B, C, and D, and Class II Groups E, F and G Hazardous Locations
    • Class I: A location where there is a danger of explosion due to the presence of a flammable gas or vapor
    • Class II: A location where there is a danger of explosion due to the presence of a flammable dust
UL / CSA Product Markings
UL / CSA Product Markings

Regulatory Requirements

instrument
  • The safest course of action is to expose the sensors to known concentration test gas before each day