Troubleshooting

4.1 Troubleshooting

PROBLEM DESCRIPTION REACTION
ERROR TEMP Temperature is below
-40°F (-40°C) or
above 167°F (75°C)
Return to normal temperature range
and recalibrate.
Contact MSA
ERROR EE EEPROM Memory error Contact MSA
ERROR PRG Flash Memory error Contact MSA
ERROR RAM RAM Memory error Contact MSA
ERROR UNK Unknown error Contact MSA

LOW
BATTERY
Battery warning repeats
every 15 seconds
Remove from service as soon as
possible and recharge battery

BATTERY
ALARM
Battery is completely
discharged
Instrument is no longer sensing
gas. Remove from service
and recharge battery
ERROR CHARGE Charge error Instrument must be between
10 and 35°C to charge.
Contact MSA if problem persists
SENSOR ERROR Missing Sensor Verify sensor is properly installed
and recalibrate.
Contact MSA if problem persists
Instrument does not turn ON Low battery Charge instrument
Sensor warning Sensor is near its end of life
& Sensor alarm Sensor has reached end of life
and cannot be calibrated.
Replace sensor and recalibrate