PROBLEM |
DESCRIPTION |
REACTION |
ERROR TEMP |
Temperature is below -40°F (-40°C) or above 167°F (75°C) |
Return to normal temperature range and recalibrate. Contact MSA |
ERROR EE |
EEPROM Memory error |
Contact MSA |
ERROR PRG |
Flash Memory error |
Contact MSA |
ERROR RAM |
RAM Memory error |
Contact MSA |
ERROR UNK |
Unknown error |
Contact MSA |
LOW BATTERY |
Battery warning repeats every 15 seconds |
Remove from service as soon as possible and recharge battery |
BATTERY ALARM |
Battery is completely discharged |
Instrument is no longer sensing gas. Remove from service and recharge battery |
ERROR CHARGE |
Charge error |
Instrument must be between 10 and 35°C to charge. Contact MSA if problem persists |
SENSOR ERROR |
Missing Sensor |
Verify sensor is properly installed and recalibrate. Contact MSA if problem persists |
Instrument does not turn ON |
Low battery |
Charge instrument |
|
Sensor warning |
Sensor is near its end of life |
& |
Sensor alarm |
Sensor has reached end of life and cannot be calibrated. Replace sensor and recalibrate |