1-866-4 JJSTECH(1-866-455-7832)

Honeywell Analytics IR-148 Infrared Gas Monitor, Eight Point, C5F8 (0-100 ppm) Rack Mount - 1324-0811

Item #:
1324-0811
Brand:
Honeywell Analytics
Honeywell Analytics IR-148 Infrared Gas Monitor, Eight Point, C5F8 (0-100 ppm) Rack Mount - 1324-0811
$29,965.00
Qty
Send This Page to a Friend

Ask an Expert

Our tech experts are ready to help!

1-866-4 JJSTECH (1-866-455-7832)

Email Us
Share this product:
Honeywell Analytics IR-148 Infrared Gas Monitor, Eight Point, C5F8 (0-100 ppm) Rack Mount - 1324-0811

Honeywell Analytics IR-148 Infrared Gas Monitor, Eight Point, C5F8 (0-100 ppm) Rack Mount - 1324-0811

$29,965.00

1324-0811 IR-148 Infrared Gas Monitor, Eight Point, C5F8 (0-100 ppm) Rack Mount by Honeywell Analytics

The Model IR-148 is an infrared gas monitor that detects toxic and combustible gases as well as solvents. It uses infrared sensing technology, allowing accurate measurement of target gas vapors with minimum interference from other vapors. The Model IR-148 is supplied as a wall mount unit configured for 1, 4 or 8 sampling points measuring down to 1 part per million (ppm). It has two available analog outputs, each which is software selectable between 10% and 100% of full scale: 4-20mA, isolated, current sourcing, and 0-10 VDC. Multiple gas calibrations are available.

Additional Features and Benefits:

  • Infrared sensing technology
  • Accurate measurement of target gas vapors with minimum interference
  • Optional relay module can be configured for 1, 2 or 3 alarm levels
  • 4-20mA

Quick Specs:

  • Response: Updated reading every 7 seconds
  • Operating temperature: 0-50°C, 32-122°F
  • Relative humidity: 0-99% non-condensing (no effect on reading)
General Specification
Typical Performance characteristics subject to change, depending on gas to be monitored and full-scale range.
For 0-1000ppm Range Accuracy: 0-100ppm ±2ppm; 100-1000ppm ±10% reading
Linearity: 0-100ppm linear, 100-1000 ±2% of full scale
Sensitivity: 2ppm Resolution: 1ppm
Note Specifications for other ranges dependant on application. Consult factory.
Reproducibility ±2ppm over 12 months at specified operating conditions
Response Updated reading every 7 seconds
Operating Temperature 0-50°C, 32-122°F
Temperature Effect ±0.3% per °C of reading
Relative Humidity 0-99% non-condensing - no effect on reading
Sample Flow Rate 1.5 liter/minute
Maximum Total Tubing Length 150f with 1 /8" ID
500f with 3 /16" ID
Operating Power Requirements 120VAC ± 10% at 0.56A, or 240VAC; ±10% at 0.3A
Alarm Relays 3 relays @ 8A resistive
Approvals CE approval: EN55011 & EN 50082-2/EN6100-4-2
Outputs Analog 4-20mA sourcing (dependent on configuration) RS23C (standard), RS422 or RS485 (optional module)
Physical Enclosure Information
NEMA 4 Dimensions: 460 x 410 x 180mm (18" x 16" x 7") (H x W x D)
Weight: 19kg (40lbs)
19" Rack Mount Dimensions: 180 x 450 x 390mm (7" x 175 /8" x 151 /4") (H x W x D)
Weight: 9kg (19lbs)

Questions & Answers