- All Products
- Shop by Brands
- BW Technologies
- BW Technologies
- BW Technologies Monitors
- Models
- BW Clip
- BW Clip4
- GasAlertClip Extreme
- GasAlert Extreme
- GasAlertMicroClip XT
- GasAlertMicroClip XL
- GasAlertMicroClip X3
- GasAlertQuattro
- GasAlertMax XT II
- GasAlertMicro 5 Series
- SamplerPak
- IntelliDox Docking System
- MicroDock II
- BW Ultra
- BW Solo
- Gas Monitors
- Gas Monitor Manufacturers
- BW Technologies
- Bacharach
- Biosystems
- Crowcon
- Gas Clip Technologies
- GfG Instruments
- GMI/Detcon
- Honeywell Analytics
- Industrial Scientific
- MSA
- RAE Instruments
- RKI Instruments
- Calibration Gas
- Combustion Analyzers
- Water Quality
- Reliability Instruments
- Refractometers
- HVAC
- Weather
- Temperature
- Laboratory
- Sound
- IAQ
- Electrical
- All Categories
Modern Water EnviroGard DDT Soil Kit - 7310000
Item #:
7310000
Brand:
Modern Water
Share this product:
7310000 EnviroGard DDT Soil Kit by Modern Water
Envirogard® is a rapid field enzyme immunoassay method for the analysis of soil remediation, assessment and industrial testing. It uses a coated tube technology to detect the analyte of concern.
Features:
- Rapid and simple field testing procedures
- Semi-quantitative data results at set calibrator (action) levels
- Test up to 17 samples at once for increased cost efficiency
- Results available in less than 30 minutes
- Coated tube immunoassay
- EPA SW-846 Method #4042
Test result type
- Semi-quantitative or qualitative data
Samples per kit
- Up to 17 soil sample test results can be obtained per kit, depending upon the number of runs and action levels used.
- Typical applications use two or three calibrators and one negative control.
Assay range
- Semi-quantitative results (greater or less than) at calibrator (action) levels of 0.2, 1.0 and 10.0 ppm DDT
- Minimum detection level (MDL) is 0.2 ppm DDT.
- No maximum screening level as extracts can be diluted into test range
Sample preparation
- Soil samples require prior extraction using the MWI Sample Extraction Kit (sold separately).
- The Sample Extraction Kit provides material for 12 soil sample extractions with methanol.
Sampling time
- Soil extraction time is typically 2 minutes per sample plus assay run time of approximately 30 minutes.
Specificity
The EnviroGard DDT immunoassay test does not differentiate between DDT and other organochlorines. The table below shows compounds at the minimum detection limit (MDL) -an approximate concentration required to yield a positive result atthe lowest standard. The IC50 is the concentration required to inhibit one half of the color produced by the negative control. It is also used to calculate cross-reactivity values to similar compounds.
Basic Test Procedure
- Extract the soil sample using the MWI Sample Extraction Kit
- Pipet DDT assay diluent into the antibody coated tubes, then add either DDT calibrator or sample extract to the tubes along with DDT enzyme conjugate.
- Wait 15 minutes, then decant and wash the tubes.
- Add substrate.
- Wait for 10 minutes, add stop solution
- Read the absorbance using a photometer, and interpret the results. (Refer to DDT Soil Test Kit Instruction Guide procedures.)
Test kit components
- 20 antibody coated tubes
- Enzyme conjugate, colour developing and colour stop solution reagents
- Calibrators for 0.2, 1.0 and 10.0 ppm DDT.
- Negative control calibrator
- Test tube rack and pipet tips.
- Test kit instructions
Storage & precautions
- Shelf life is typically one year from date of manufacture, with specific kit expiration date information provided on product packaging
- Components must be stored at 39° to 46°F (4° to 8°C) when not in use
- Store at ambient temperature 64° to 81°F (18° to 27°C) is acceptable for day of use
- Kits must be brought to 64° to 81°F (18° to 27°C) before use
- Do not expose colour solution to direct sunlight
- Portable spectrophotometer battery should be charged prior to use
Required test materials
- EnviroGard® DDT Soil Kit (7310000)
- Sample extraction kit (soil)(7310000EB)
Required test equipment
- Field soil lab (Accessory kit) (6050400) which includes: Repeating pipette*, Adjustable pipette*, Balance*, Repeating pipette tips*, Timer*
- Differential spectrophotometer - 110V (6000001), 220V (6000002)
- Field lab system rental (includes items marked with * above) (6997020)
Other recommended materials
- Latex gloves
- Absorbent paper for blotting test tubes
- Liquid and solid waste containers
- Calculator
Contaminant | MDL | IC50 |
p,p’-DDT (kit calibrator) | 0.2 | 1.25 |
p,p’-DDD | 0.05 | 0.3 |
p,p’-DDE | 0.6 | 3.6 |
o,p’-DDT | 14.9 | 93 |
o,p’-DDD | 1.76 | 11 |
o,p’-DDE | 14.9 | 93 |
DDA | 0.01 | 0.04 |
Chloropropylate | 0.01 | 0.08 |
Chlorobenzilate | 0.06 | 0.35 |
Dicofol | 0.3 | 2 |
Tetradifon | 2.2 | 14 |